Buffalo Filter ViroVac™ Operator’s Manual: Page 11

• FILTER LIFE INDICATOR

(Membrane Control Panel)

The filter life indicator on the membrane control panel provides a visual indication of the status of

the life of the filter in use. The filter life indicator for the ViroVac™ Smoke Evacuation System will

automatically adjust according to the flow setting selected.

Low Flow Setting = 35 hours of filter life

Medium Flow Setting = 24 hours of filter life

High Flow Setting = 18 hours of filter life

The ViroVac™ Filter Life Indicator is factory set. All filter life timing is automatic.

Reading the Filter Life Indicator:

Install an unused ViroSafe® Filter into the system per the installation instructions contained in

this operator’s manual. When the system is turned on, the filter life indicator with light up the

leftmost GREEN LED, indicating 100% filter life. The indicator will progress thru subsequent

GREEN LEDs, to an AMBER LED as time elapses and begin flashing RED to indicate that the

filter has expended its useful life and requires replacement.

When the maximum filter life is consumed and the smoke evacuator is not powered off for

greater than six (6) hours or the main power is disconnected – a new filter is required to acti-

vate the smoke evacuator and make operational.

• FUSES

(circuit board)

Two 10 AMP fuses (8 AMP for 220/240 ViroVac™ systems) are located on the circuit board within

the housing of the system. It electrically protects both the system and the operator from damage

or injury. If the system is overheated or if there is an electrical surge in the electrical system, fuses

will break and the system will not operate.

When the Service light illuminates, please contact BUFFALO FILTER Customer Service for

system service instructions.