In addition, please compare the accessories you receive with the standard accessories list below. If

an item is missing, please notify BUFFALO FILTER Customer Service.

Standard Accessories:

• Operator’s Manual

• Power Cord

• Pneumatic Footswitch

Please contact BUFFALO FILTER Customer Service to purchase the following accessories:

• Replacement Filters

• EZ Link Remote Activation Device

• Hoses, Tubing, Laparoscopic Kits, Adapters, Wands & Other Accessories

1.3 Operational Information

internationally:

100/120 VAC

50/60 Hz, 220/240 VAC 50/60 Hz

2. Type of protection against electrical shock (UL 60601-1, Clause 5.1): Class I

3. Degree of protection against electric shock (UL 60601-1, Clause 5.2); Type CF Applied Part

4. Degree of protection against ingress of water (UL 60601-1, Clause 5.3): IPX1

5. Method of sterilization or disinfection recommended by BUFFALO FILTER (UL 60601-1,

Clause 5.4):

Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution

or soapy water. Wipe dry with a clean cloth. Do not steam sterilize.

with oxygen or nitrous oxide (UL 60601-1, Clause 5.5): Not Suitable

7. Mode of operation (UL 60601-1, Clause 5.6): Continuous

8. Upon request, BUFFALO FILTER will provide the following:

Service and Repair Instructions, including Circuit Diagrams and Parts List

cian as follows:

100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo), (F1, F2)

220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo), (F1, F2)