Buffalo Filter ViroVac™ Operator’s Manual: Page 21

3.1

General Maintenance Information

This section contains information for ordinary upkeep of the BUFFALO FILTER ViroVac™ Smoke

Evacuation System. While the system has been designed and manufactured to high industry

standards, it is recommended that periodic inspection and performance testing be performed by

a qualified Biomedical Technician to ensure continued safe and effective operation.

3.2 Cleaning

Unplug Unit. Wipe unit with a damp cloth containing mild disinfectant solution or soapy water.

Wipe dry with a clean cloth. Do not steam sterilize.

3.3

Periodic Inspection

The ViroVac™ Smoke Evacuation System should be visually inspected at least every year. This

inspection should include checks for:

• Damage to the power cord.

• Damage to the power plug or power inlet module.

• Proper mating, cleanliness and absence of damage to the filter inlet.

• Obvious external or internal damage to the system.

3.4

Troubleshooting the System – see below.

Maintenance

Section 3.0

PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION

1. Smoke Evacuation System

is ON but suction is minimal

or none.

2. Smoke Evacuation System

does not function even

though suction ON button is

depressed.

1. Filter is not seated

completely.

2. Filter is clogged.

3. Vacuum hose or tube is

clogged.

4. Motor/blower is obstructed.

1. Not plugged into an

electrical outlet.

2. Fuses are blown.

3. Electronic system failure.

4. Filter life has expired or

invalid filter inserted.

1. Re-install ViroSafe Filter, press firmly into place

and fully engage clip.

2. Replace filter with a genuine Buffalo Filter

ViroSafe filter.

3. Replace vacuum hose or tube with genuine

Buffalo Filter products.

4. Call BioMed or BUFFALO FILTER Technical

Services at 1.800.343.2324 or 716.835.7000.

1. Check power outlet and connection to rear or

side panel of the machine.

2./3. Call BioMed or BUFFALO FILTER Technical

Services at 1.800.343.2324 or 716.835.7000.

4. Replace filter with a genuine Buffalo Filter

ViroSafe filter.