92 Calibration Using Converter Signal Source

9.2.1

0 % Flow Kate Calibration

■ To check the ?:ero point of flow measurement:

Key operation

Display example

Description

Sin

N1:FLOW CAL OK

0.0 K

Press [SRT] first to Stan the items selection sequence

and select N1: FLOW CAL0% from among the
configuration iteras using [ ► ) and [

^

] keys. Then

press [SET] again to go into the calibration mode

and calcúlale and display zero point using a

simulation signal.

sur

N1:FLOW CAL OK

Pressing [SET], the system returns lo the items

selection sequence.

To change the zero and span of the converter:

Key operation

Display example

Description

SET

N1:FLOW CAL OK

0. 1 K

Press [SH'I’] first to start the items selection sequence

and select N1: RX>W CAL0% from among the
setting items using [ ► J and [

a

] keys. Then press

[SET] again to go into the calibration mode and

calculate and display zero point using a simulation

signal.

ADJUST READY

0.1

%

Pressing [ ► ]/'ADLTUiir READY” appears as

shown left and the system goes into a calibration

ready condition, (Note 1)

N1:FLOW CAL OK

* CAL. OK ADJ.

Pressing [SET], ”*CAL0% ADJ.” appears as shown

left and the system staris the zero caltbraiion, The

zero calibration takes about 3 to 6 seconds. (Note 2)

SET

N1:FLOW CAL OK

3. 0

%

Newly calibrated zero point appears.

SET

N1:FLOW CAL OK

Pressing [SET], the system returns to the items

selection sequence,

Notes:

h'10 caned zero calibrations press [

a

. ]. 'Jhe sy^stem reí urns to the puint where the zero point

display appCctrs.

2.Calibration lime depends on the excitation Irequency (24

y \ ? :

3 sec* 12 llz and 6

U z :

fi sec).

-

104

-