DIP SWITCH FUNCTION SUMMARY

S1-1 S1-2

OPERATION/SETUP

OFF

OFF

Normal operation (this is the normal setting of these switches)

ON

OFF

Access configuration mode

OFF

ON

Access analog output (4-20mA) test and calibration

ON

ON

Access scale diagnostic and advanced calibration parameters

Note: The calibration DIP switch will override the OPERATION/SETUP switches.

S1-3 WEIGH

UNITS

OFF

Pound (lb) scaling

ON

Kilogram (kg) scaling


S1-4 CALIBRATION

OFF

Normal operation (this is the normal setting for this switch)

ON

Calibrate scale w/ dead zero





MESSAGES

DISPLAY

EI-1000

DISPLAY

EI-2000

DESCRIPTION

nO O

nO O

No zero – The scale cannot be zeroed since the weight exceeds
the value set by the “Push to Zero/Auto-zero PTZ/AZ maximum
range” parameter (Px14).

FACTRY FACTORY Loading factory default, which is initiated by the user
AUtO.Ld AUtO.LOAd Automatically loading default configuration & calibration

Add Lb

Add KG

Add

Add more weight to the scale before trying to adjust gross weight

A OL

A OL

Scale input positive overload

-A OL

-A OL

Scale input negative overload

FAIL.AD FAIL.AD *Channel 1 & 2 A/D hardware failure.

IIC.Err IIC.Err

*IIC bus error

EE.Err EE.Err

*EEPROM device was not found

* Hardware failure: Return the EI-1000/EI-2000 to factory for service.