3 cobalt photography mode, 4 fluorescein angiography mode, Ge 21 – Canon CX-1 User Manual

Page 33

Advertising
background image

5. Basic Photography

-21-

5.3.3 Cobalt Photography Mode

In this mode, photography using the exciter filter for fluorescein angiography is possible.

1

Enter the study information into the computer.

2

Select the cobalt photography mode.

Press the COBALT switch.
Or select the photography mode with the control
software.
The COBALT lamp turns on. COBALT is displayed on
the observation monitor.

3

Perform the procedure starting from step 3 in 5.3.1 Color Photography Mode (see page 14).

5.3.4 Fluorescein Angiography Mode

If taking multiple photographs, a mydriatic must be used.

1

Enter the study information into the computer.

2

Select the fluorescein angiography mode.

Press the FLUO switch.
Or select the photography mode with the control
software.
The FLUO lamp turns on. FLUO is displayed on the
observation monitor.

3

Disinfect the forehead rest and replace the chin
rest paper.

4

Have the patient sit down.

5

Align the height of the patient’s eye.

6

Move the stage position.

7

Align the retinal camera to the eye to be photographed.

8

Adjust brightness in the field of view for observation.

Advertising