5. Basic Photography

-25-

5

Align the height of the patient’s eye.

Press the CHIN REST switches to adjust the chin rest up or down so that the patient’s eye is aligned
with the height adjustment mark.

6

Move the stage position.

If the stage is locked at the operator’s side, press the
stage lock button to unlock the stage.
Hold the operation lever, pull it towards you, and then
move the stage toward the eye to be photographed.

7

Align the retinal camera to the eye to be photographed.

Slowly bring the retinal camera closer to the patient
eye.

CAUTION

Do not put your fingers or hands between the sliding part and base of the
stage unit or on the panning and tilting sliding parts.
Similarly, do not allow patients to put their hands in these places.
Otherwise, your or their hands or fingers may be pinched, possibly
resulting in injury.

CAUTION

When adjusting the position of the retinal camera in the front-back
direction, slowly bring the retinal camera closer to the patient while looking
at the patient’s eye from the side.
You may injure the patient’s eye if the objective lens makes contact with it.