8. Main Specifications and Performance

-56-

Electromagnetic Immunity

The CX-1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the CX-1 should assure that it is used in such an
environment.

Immunity Test

IEC 60601 Test

Level

Compliance

Level

Electromagnetic Environment –

Guidance

Electrostatic
discharge (ESD)

IEC 61000-4-2

±(2, 4, 6) kV
contact

±(2, 4, 8) kV air

±(2, 4, 6) kV
contact

±(2, 4, 8) kV air

Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.

Electrical fast
transient/burst

IEC 61000-4-4

±2 kV for power
supply lines

±1 kV for input/
output lines

±2 kV for power
supply lines

±1 kV for input/
output lines

Mains power quality should be
that of a typical commercial or
hospital environment.

Surge

IEC 61000-4-5

±1 kV
differential
mode

±2 kV common
mode

±1 kV
differential
mode

±2 kV common
mode

Mains power quality should be
that of a typical commercial or
hospital environment.

Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines

IEC 61000-4-11

<5% U

T

(>95% dip in
U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in
U

T

) for 5 sec

<5% U

T

(>95% dip in
U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95% dip in
U

T

) for 5 sec

Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the CX-1 requires continued
operation during power mains
interruptions, it is recommended
that the CX-1 be powered from an
uninterruptible power supply.

Power frequency
(50/60Hz)
magnetic field

IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.