18.5 Alibi memory

The weighing scale software can save up to 100 000 measurements. Measurements
are saved automatically each time the PRINT-key is pressed. In addition to
measurements other data is also saved:

• Date of measurement

• Time of measurement

• Measuring result (weight)

• Tare value

• Logged-in user

• Name of weighed product


• Measurement one will be deleted as soon as measurement 100 001 is

saved.

• Measurements saved to the alibi memory cannot be deleted.

1

Number of measurement:
Format: b4.5n N=number of measurement

2

Date of measurement

3

Time of measurement


Each can be invoked as described above (chap. 18.4).
If menu item b4.5.1.7 is selected and a printer connected, data can be printed by
pressing PRINT. (Dependent on settings selected in GLP report chap. 15).