2 remote control instructions – KERN & SOHN AEJ 200-4CM User Manual

Page 94

Advertising
background image

16.4.2 Remote control instructions

Feedback by weighing scale after sending remote control command:

XX_A

CR LF

Instruction accepted; will be executed

XX_I

CR LF

Instruction received; impossible to carry out

XX_^

CR LF

Command accepted but max range exceeded

XX_v

CR LF

Command accepted but min range exceeded

XX_E

CR LF

Error during execution, timeout for stable weighing value
exceeded

XX

Name of command

1 – 3

4

5

6

7

8 – 16

17

18 - 20

21

22

C

om

m

a

nd

B

lank

S

ta

b

ili

ty

s

y

m

bol

B

lank

C

har

ac

ter

W

ei

gh

t

B

lank

Un

it

CR

LF



Instruction:

1-3

Character

Stabilisation
symbol

[blank space]

If stable

[?]

If stable

[^

]

If error (Command + exceeded)

[v]

If error (Command – exceeded)

Character

[blank space]

For positive and negative values

Weight

9 digits

With right justifications

Unit

3 digits

With left justification


94

AEJ-C/AES-C/PLJ-C-BA-e-1412

Advertising
This manual is related to the following products: