6.8 Non-verifiable models:

 Switch on the balance and during the selftest press the

and the

button at the same time.

„UnLoad“ is displayed shortly.

 Ensure that there are no objects on the weighing pan.

 „LoAd“ is displayed

 Place adjustment weight, the stability display appears

 After the adjustment the balance will carry out a self-test.

„PASS“ will be displayed shortly. Remove adjusting weight
during selftest, the appliance will return into weighing mode
automatically.
An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.

6.9 Verifiable models:


 Switch on balance and press the adjustment switch at the

same time.

 After that, during the balance’s selftest, press

and

at the same time.

„UnLoad“ is displayed.

 Acknowledge with

. Ensure that there are no objects on

the weighing pan.

 The currently set adjustment weight will be displayed.

 To change by using the navigation buttons (see chap. 8)

select the desired setting, the active digit is flashing.

 Acknowledge with

. „LoAd“ is displayed.

18

GAB-P-BA-e-1411