9 manual totalizing – KERN & SOHN GAB 30K-3PM User Manual

Page 28

Advertising
background image

7.9 Manual totalizing

This function is used to add the individual weighing values to the summation memory
and to provide a printout via the installed printer, when the stability display appears

and the

button is pressed.

(in order to set this function see chap. 8.2 „ACC on“)

 Place material to be weighed A

Wait until stability display appears

(example)

 Press

„ACC 1“ followed by the weight value will appear. The weight
value is added into the summation memory and will be

printed out when the

button is pressed

 Remove material to be weighed A

Further material to be weighed can only be added after the
balance has returned to zero.

 Place material to be weighed B

Wait until stability display appears

(example)

 Press

„ACC 2“ followed by the weight value will appear. The weight
value is added into the summation memory and will be

printed out when the

button is pressed

 Remove material to be weighed B

Further material to be weighed can only be added after the
display has returned to zero.

 Add more material to be weighed as described before.

Please note that the balance must be unloaded between the
individual weighing procedures.

 You can repeat this process until the capacity of the scales

is exhausted.

28

GAB-P-BA-e-1411

Advertising
This manual is related to the following products: