CAUTION!

• To prevent electrostatic charge, connect the balance to earth.
• The contact slots for the wires are marked with the icon „ “.


• Prior to transport, prepare a PE conductor for electrostatic connection to earth of

the display unit.

• The PE conductor should end in a loop that is screwed to the display unit with the

help of nuts (M4).

• To ensure that the loop is fixed permanently and in contact with the display unit,

use a lock washer.

• The PE conductor should be of yellow and green colour and provide a sectional

surface of 4mm².

The diagram below shows the balance’s connection to earth:


Place the platform balance and the display unit on the intended installation site in the
ex-area and connect it to earth.

Transport Securing

Weighing pan

Safe area

Ex-protected area