6.6 Linearization
Linearity shows the greatest deviation of a weight display on the scale to the value of

the respective test weight according to plus and minus over the entire weighing

range. If linearity deviation is discovered during a testing instrument control, you can

improve this by means of linearization.

Carrying out linearization is restricted to specialist staff possessing

well acquainted with the workings of weighing scales.

The test weights to be used must be adapted to the weighing scale’s

specifications; see chapter “testing instruments control”.

Observe stable environmental conditions. Stabilisation requires a

certain warm-up time.

Do not remove the adjustment weight during linearization in step

LAOD 1

to LOAD 4, merely increase it instead. Conversely do not

remove the adjustment weight during step LAOD 4 to LOAD 1,

merely increase it instead.

After successful linearisation you will have to carry out calibration;

see chapter “testing instruments control”.

Tab. 1: Adjustment weights“LOAD1 – LOAD4”

MAX

LOAD 1

LOAD 2

LOAD 3

LOAD 4

3kg

0.5kg

1kg

2kg

3kg

6kg

1kg

2kg

4kg

6kg

15kg

3kg

5kg

10kg

15kg

30kg

5kg

10kg

20kg

30kg

60 kg

10kg

20kg

40kg

60kg

120 kg

30kg

60kg

60kg

150kg

300 kg

50kg

100kg

200kg

300kg

600 kg

100kg

200kg

400kg

600kg

1.5 t

3000kg

5000kg

1000kg

1500kg

3 t

5000kg

1000kg

2000kg

3000kg