7.5 Weighing with tare

Deposit weighing vessel. After successful standstill control press the

button.

Zero display and the indicator [] next to

T AR E

appear. The weight of the

container is now internally saved.

Weigh the material, the net weight will be indicated.

After removing the weighing container, the weight of the weighing container

appears as negative display.

The tare procedure can be repeated as many times as necessary, for example

with initial weighing of several components for a mix (add-on weighing). The limit

is reached when the total weighing range capacity is full.

To delete the tare value, remove load from weighing plate and press

.

7.6 Counting
During piece counting parts can either be counted into a container or out of a

container. To count a greater number of parts the average weight per part has to be

determined with a small quantity (reference quantity). The larger the reference

quantity, the higher the counting exactness.

High reference must be selected for small parts or parts with considerably different

sizes.

The average piece weight can only be determined by stable weighing

values.

If weighing values are under zero, the piece counter display shows a

negative number of items.

The message light appearing on the display indicates that load falls below

minimum weight value.

Delete incorrect entries by pressing

.

The accuracy of an average item weight can be improved at any time

during additional counting processes. For this purpose add additional

items and press

. After the reference optimization sounds a signal

tone. As the additional pieces increase the base for the calculation, the

reference also becomes more exact.