Carry out adjustment:

Confirm menu setting nonLin with

.

Ensure that there are no objects on the weighing pan.

LoAd will be displayed after standstill control has been carried out.

Put the required adjustment weight carefully in the centre of the weighing pan.

After successful adjustment, the weighing scales will carry out a self test. During

this self test remove the adjustment weight and the weighing scales will

automatically return to weighing mode.

An adjusting error or incorrect adjusting weight will be indicated by the error

message; repeat adjustment procedure.