4 mains connection, 5 adjustment – KERN & SOHN IFS 300K-3 User Manual

Page 13

Advertising
background image

KFS-T-BA_IA-e-1213

13

6.4 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be

the same as the local voltage.

Only use original KERN mains adapters. Using other makes requires consent by

KERN.

6.5 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,

each display unit with connected weighing plate must be coordinated - in compliance

with the underlying physical weighing principle - to the existing acceleration due to

gravity at its place of location (only if the weighing system has not already been

adjusted to the location in the factory). This adjustment process must be carried out

for the first commissioning, after each change of location as well as in case of

fluctuating environment temperature. To receive accurate measuring values it is also

recommended to adjust the display unit periodically in weighing operation.

Provide adjustment weight.

The required adjustment weight depends on the capacity of the

weighing system. Carry out adjustment as near as possible to the

scale’s maximum weight. Information about test weights you will find in

the internet under: http://www.kern-sohn.com

Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.