12 Installing display unit / weighing bridge

Installation / configuration of the weighing system must be carried out by a well

acquainted specialist with the workings of weighing balances.

12.1 Technical data

Supply voltage:

5 V/150mA

Sensitivity

2-3 mV/V

Resistance parameter

80 - 100 Ω, max 4 items per 350 Ω load cell

12.2 Weighing system design
The display unit is suitable for connection to any analogue platform in compliance

with the required specifications.

The following data must be established before selecting a weighing cell:

Weighing balance capacity

This usually corresponds to the heaviest load to be weighed.

Preload

This corresponds to the total weight of all parts that are to be placed on the

weighing cell such as upper part of platform, weighing pan etc.

Total zero setting range

This is composed of the start-up zero setting range (± 2%) and the zero

setting range available to the user via the ZERO-key (2%). The total zero

setting range equals therefore 4 % of the scale’s capacity.

The addition of weighing scales capacity, preload and the total zero setting

range give the required capacity for the weighing cell.

To avoid overloading of the weighing cell, include an additional safety

margin.

Smallest desired display division