Procedure when adjusting:

Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization.

Switch on the balance using the

key

Press

key and keep it pressed, after the acoustic signal appears in the display for

short time „CAL“. After that the exact size appears flashing in the display
(chapter.7.4) of the adjustment weight.
Now set the adjusting weight in the centre of the weighing plate.

Now press the

key. Short time later appears „CAL F“, then return automatically to

the normal weighing mode. In the display there appears the value of the adjustment
weight.
An error during adjustment or the use of an incorrect adjusting weight will result in an
error message „CAL E“. Repeat adjustment.

Keep the adjustment close to the balance. Daily control of the weighing exactness is