KERN & SOHN DE User Manual

Page 18

Advertising
background image


DE-BA-e-1255

18

6.3

Taring

Switch-on the balance using the

key and wait for the

„0“ display.

Put the tare vessel on the weighing plate and press the

key. The balance

display goes to „0“. The weight of the container is now internally saved.

If after finishing the weighing process the

key, is pressed

again, „0“ appears

anew in the display.

The taring process can be repeated any number of times, e.g. when adding several
components for a mixture (adding).
The limit is reached when the whole weighing range is exhausted.
After removing the taring container the total weight is displayed as negative display.

6.4 PRE-Tare function


Using this function the weight of a tare vessel can be stored.
This value also remains saved if the balance meanwhile has been switched off and
switched on again.

Switch-on balance using the

key and wait for the „0“ display.

Put tare vessel on the weighing plate and press the

key 6 times until „PtArE“

flashes on the display. By actuating the

key, the current weight on the weighing

plate is saved as PRE-Tare weight.

To switch off this function, unload the weighing plate and press the

key 6 times,

until „PtArE“ flashes on the display. Then press the

key. The stored PRE-Tare

weight is deleted.

6.5 Plus/minus weighings

For example unit weight control, fabrication control etc.

Switch-on the balance using the

key and wait for the

„0“ display.

Put the no

minal weight on the weighing plate and tare to „0“ using the

key.

Remove the nominal weight.
Put the test objects subsequently on the weighing plate, the respective deviation
from the nominal weight is displayed with the re

spective sign to „+“ and „-“.

According to the same procedure also packages with the same weight can be
produced, referring to a nominal weight.

Back to weighing mode by pressing the

key.

Advertising