KERN & SOHN DE User Manual

Page 34

Advertising
background image


DE-BA-e-1255

34

8.3.4 rE Cr

The remote control commands s/w/t are sent from the remote control unit to the
balance as ASCII code. After the balance having received the s/w/t commands, it will
send the following data.

Take into account that the following remote control commands must be sent
without a subsequent CR LF.


s

Function:

Stable weighing value for the weight is sent via the RS232
interface

w

Function:

Weighing value for the weight (stable or unstable) is sent
via the RS232 interface

t

Function:

No data are sent, the balance carries out the tare function.


h. Format for stable values for weight/quantity/percentage
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B U

1

U

2

U

3

CR

LF


i. Format in case of fault
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

B B B B B B B B B B B E

r

r

o

r

CR

LF


j. Format for unstable values for weight/quantity/percentage
1

2

3

4

5

6

7

8

9

10 11 12

13 14 15 16 17

18

M

S

N

1

N

2

N

3

N

4

N

5

N

6

N

7

N

8

N

9

N

10

B B B B CR

LF


Symbols

M

Blank or M

S

Blank or minus sign (-)

N

1

… N

10

10 numeric ASCII codes for weight values including decimal places or

blanks

U

1

… U

3

3 ASCII codes for weighing unit pcs. / % / or blank

B

Blank

E, o, r

ASCII code or “E, o, r”

CR

Carriage Return

LF

Line Feed

Advertising