9 service documentation, 1 overview, setting specification, tolerances – KERN & SOHN KFU 1500V30M User Manual

Page 13

Advertising
background image

KFU-IA-e-1211

13

9 Service documentation

• This chapter is only provided for a specialist of balances !

• The weighing bridges are carried out in DMS sensor technology, at

every corner there is a DMS weighing cell.

• The analogue-digital conversion takes place in the display unit. There

are also saved all balance and country specific data.

9.1 Overview, setting specification, tolerances

Test and setting specification:

Capacity

600 kg

1500 kg

Readability

200 g

500 g

Min

4 kg

10 kg

Max

600 kg

1500 kg

1/3 corner load

200 kg

500 kg

Tolerance

200 g

500 g

Verification data and tolerances acc. to OIML:


-1

-0,75

-0,5

-0,25

0

0,25

0,5

0,75

1

0

500

1000

1500

[kg]

[g]

-0,4

-0,3

-0,2

-0,1

0

0,1

0,2

0,3

0,4

0

100

200

300

400

500

600

[kg]

[g]

600 kg

1500 kg

Advertising