KERN & SOHN VFB 2T-3 User Manual

Page 12

Advertising
background image

VFB-BA-e-1310

12

Non-optimal positioning of load

Optimal positioning of load

 Temperature range: The maximum deviation for 0.1% of the weighed load is

between -10 and +40°C. Outside this temperature range deviations of up to
0.3% may occur.

 Avoid jarring during weighing

 Protect the weighing system against high humidity, vapours and dust

 Do not expose the device to extreme dampness for longer periods of time.

Non-permitted condensation (condensation of air humidity on the appliance)
may occur if a cold appliance is taken to a considerably warmer environment.
In this case, acclimatize the disconnected appliance for ca. 2 hours at room
temperature.

 Major display deviations (incorrect weighing results) may be experienced

should electromagnetic fields (e.g. due to mobile phones or radio equipment),
static electricity accumulations or instable power supply occur. Change
location or remove source of interference.























Pallet

Load