22

Invoke menu item

, see chapter 9.1 “How to invoke a menu“.

How to carry out adjustments

 Acknowledge with

.

Ensure that there are no objects on the weighing plate.

 Wait for stability display, then press

. The currently set

adjustment weight will be displayed.

(example)

 To change by using the navigation buttons (see chap.

5.10.1) select the desired setting, the active digit is flashing.

 Acknowledge by

.

 Carefully place adjusting weight in the centre of the weighing

plate. Wait for stability display, then press

.

 After the adjustment the balance will carry out a self-test. Err

4 will appear briefly, followed by weighing value. The system
is now in weighing mode.

Remove weight, „0“ is displayed.

An adjusting error or incorrect adjusting weight will be
indicated by the error message; repeat adjustment
procedure.