5.7.1 Data-Hold function

Switch on the balance, keep the

HOLD-key pressed until the current setting „Hx“

(H1 – H6) appears.

Press

ON/OFF button repeatedly until the desired setting „H1-H4“ is displayed.

Confirm setting with the

HOLD button.

Suspend the material to be weighed

The weight value is fixed and displayed depending on the setting (H1 – H4) (see
chap. 5.7), represented by the [

] in the upper left.

5.7.2 Animal weighing function
This function is suitable for busy weighing procedures. The result is a mean value
formed by 16 weighing values which is found out within 3 seconds.

Switch on the balance, keep the

HOLD-key pressed until the current setting „Hx“

(H1 – H6) appears.

Press

ON/OFF button repeatedly until the setting „H5“ is displayed.

Confirm setting with the

HOLD button.

Suspend the material to be weighed

Press the

HOLD-key, the display counts reverse from 3 -1.

The calculated mean value is indicated, symbolised by the [

] top left.

Before further measurements press first the

ON/OFF/TARE-key.