How to carry out adjustment:

Confirm adjustment function “Commo“ by

.

“ULoAd“ will be displayed. Make sure that no loads
apart from the lifting tackle are attached to the hook.

Wait for stability display, then press

.

Either use the displayed adjustment weight or change

by

and

, the currently enabled digit is

flashing.

To change the selected (flashing) digit, press

repeatedly until the desired value is displayed. Then

select further digits by

and amend them by

.

(example)

Confirm by

, „LoAd“ will be shown.

Attach adjustment weight. Wait for stability display,

then press

.

After successful adjustment „Pass“ will be displayed.
Afterwards the balance will carry out a self-test.
“Err19“ will appear briefly (Ignore error message), then
the balance will change automatically into weighing
mode and the total weight will be displayed.
Adjustment has now been completed successfully.

(example)

In case of an adjustment error or incorrect adjustment value the error message
“fail“ will be displayed and you must repeat the adjustment procedure.