7.2

Linearization

Linearity shows the greatest deviation of a weight display on the scale to the value of
the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a monitoring of test resources, you can
improve this by means of linearization.

• Carrying out linearization is restricted to specialist staff possessing well

acquainted with the workings of balances.

• The test weights to be used must be adapted to the balance’s

specifications; see chapter 2.9 “Testing instruments control”.

• Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.

• After successful linearization you will have to carry out calibration; see

chapter 2.9 “Testing instruments control”


Preparation:
Call up linearization function “LinE0“; see chap. 7.1

How to carry out linearization:

Confirm „LinE0“ by

.

“LoAd0“ will be displayed. Make sure that no loads apart
from the lifting tackle are attached to the hook.

Wait for stability display, then press

.

As soon as “LoAd 1“ is displayed attach first adjustment
weight (1/3 max).

Wait for stability display „STABLE“, then press

.

As soon as “LoAd 2“ is displayed attach second adjustment
weight (2/3 max).