Laboratory Equipment Pty Ltd. 2 LEVEL INCUBATOR SHAKER 2013 VER1.01

Refrigeration

Press the

Set/Confirmation

key, then press the

Increase

key to enter the password 9.

Press the

Refrigeration

key, the unit enters into refrigeration setting mode.

Refrigeration parameter can be set via the

Increase/Decrease

key. To confirm the new

setting, press the

Set/Confirmation

key.

If the unit is to be used in the temperature range of ambient-5°C to ambient+5°C, the
refrigeration parameter can be set at the value 0.2.


If the unit is to be used in the temperature range of ambient-20°C to ambient-5°C, the
refrigeration parameter can be set at the value 0.5.

If the unit is to be used in the temperature range of ambient+5°C to 60°C, the
refrigeration parameter can be set at the value 2. The refrigeration is cancelled.



6. Overtemperature Alarm

In case the gap between measured temperature and setting temperature excels ± 3°C,
the unit sends out audible and visible alarm automatically, and the yellow alarm
indicator flashes. The platform keeps on shaking while the heating or refrigeration
stops. Press the

Temperature

key will cancel the alarm for 30 minutes. Reset the

temperature value, the unit exits the alarm mode.


7. Speed Control


To set the desired shaking speed, first press the

Set/Confirmation

key, then press the

Increase

key to enter the password 3. Press the

Speed

key, the unit enters into speed

setting mode. The desired speed value can be set via the

Increase/Decrease

key. To

confirm the new setting, press the

Set/Confirmation

key. Setting of a new value

overwrites the old value in the memory.


8. Timer

When timer function is needed, first press the

Set/Confirmation

key, then press the

Increase

key to enter the password 3. Press the

Timer

key, the unit enters into timer

setting mode, the desired timer value can be set via the

Increase/Decrease

key. To

confirm the new setting, press the

Set/Confirmation

key.

By the time the remaining time reaches 0:00, the platform stops shaking, the unit sends out audible
and visible alarm for 2 minutes. Press the

Timer

key, the alarm will be muted.