10. Non-volatile Memory Function

Non-volatile memory retains parameters during a power interrupt and restarts the
incubator shaker automatically. Non-volatile Memory Function can be set by the
following steps:

1. First press the

Set/Confirmation

key, then press the

Increase

key to enter the

password 5.

2. Press the

Timer

key, the unit enters into non-volatile memory setting mode.

The LED indicator displays ON/OFF options.

3. Press the

Increase/Decrease

key to choose between ON and OFF. When ON is

selected, the non-volatile memory function is started. When Off is selected,
the memory function is canceled.



11. Technical Specifications

Item

Incubator
Shaker

Incubator
Shaker

Incubator
Shaker

Incubator
Shaker

Model

ZHWY-2102

ZHWY-2102C

ZHWY-2112B

ZHWY-2112F

Performance

Heated and Refrigerated

Heated and Refrigerated

Temperature
Range

Ambient: -20°C to 60°C

Ambient: -20°C to 60°C

Temperature
Set Division

±

0.1°C

±

0.1°C

±

0.1°C

±

0.1°C

Temperature
Uniformity

±

1°C

±

1°C

±

1°C

±

1°C

Shaking Speed

60-300 rpm
(±1rpm)

60-300 rpm
(±1rpm)

60-300 rpm
(±1rpm)

60-300 rpm
(±1rpm)

Timer

0-500 hours

0-500 hours

0-500 hours

0-500 hours

Capacity

50mlx90 or
100mlx90 or
250mlx56 or
500mlx48 or
750mlx36 or
1000mlx30

50mlx48 or
100mlx48 or
250mlx30 or
500mlx24 or
750mlx18

50mlx154 or
100mlx154 or
250mlx90 or
500mlx70 or
750mlx56 or
1000mlx36

50mlx154 or
100mlx154 or
250mlx90 or
500mlx70 or
750mlx56 or
1000mlx36

Platform Size

734x458 mm

500x350 mm

970x560 mm

940x580 mm

Overall Size

700x590x1230
mm

950x710x1250
mm

1200x820x1600
mm

1200x820x1600
mm

Electrical
Requirement

240V 50Hz

240V 50Hz

240V 50Hz

240V 50Hz

Weight

235Kg

150Kg

456Kg

486Kg

Warranty

1 year

1 year

1 year

1 year











Advertising