9. Defrosting

Operation the unit at a low temperature for a long period of time causes the buildup of
frost, which degrades the freezing performance and may disable the temperature from
being controlled at the setpoint. The defrosting function guarantees the unit to operate
smoothly and continuously in such a case. Defrosting function is set via the following
steps:

• First press the

Set/Confirmation

key, then press the

Increase

key to enter the

password 9. Press the

Refrigeration

key, the unit enters into refrigeration mode

(first stage).

• Press the

Refrigeration

key again, the unit enters into defrosting timer setting

mode (second stage). Defrosting timer value can be set via the

Increase/Decrease

key.

• Press the

Refrigeration

key once more, the unit enters into defrosting time

period setting mode. Defrosting time period can be set via the

Increase/Decrease

key.

• To confirm the new setting, press the

Set/Confirmation

key.


Defrosting Timer Setting

refers to the function by which the unit starts defrosting every given time

period. For example, if the defrosting timer value is set at 1, the unit starts defrosting every 1 hour.
If the defrost timer value is set at 2, the unit starts defrosting every 2 hour. However, if the defrost
timer value is set at 0, the unit does not start defrosting at all.

Defrosting Time Period

refers to the

time span

from the moment the unit starts defrosting till it

exits defrosting mode. For example, if the defrosting time period is set at 1, the unit will defrost
for 1 minute. If the defrosting time period is set at 2, the unit will defrost for 2 minutes.

Due to the specific application conditions, the defrosting timer value and defrosting time period
may vary greatly. However, the following values can be viewed as a starting point or general rule.

Defrosting Timer Value: 6
Defrosting Time Period: 3

If the above values are set, the unit will defrost for 3 minutes every 6 hours under refrigerated
operation. During defrosting, the water tray enclosed in the package must be taken out for fetching
the melted water.