LABEC M-MIA-A User Manual

Page 33

Advertising
background image

Ultrasonic milk analyser

6.3.2.1. Press the button Enter and without releasing it switch on the power
supply of the device, wait for the starting identification messages and release
the button after the following message appears on the display:

Release button

to start setup


After releasing the button on the display is shown:

MA Setup


followed by possible to be entered by the operator menus:


Special modes
Corrections
Settings
------------------------------
Tests
pH & Co Meter
Exit


6.3.2.2. By using buttons “up” ▲ and “down” ▼position on Corrections and
press Enter.

6.3.3. Making correction

6.3.3.1 Determining the correction mode
When starting Corrections, the following appears on the display








Corrections:
Calibration 1
Calibration 2
Calibration 3
---------------------------
Temperature
Exit

Operation manual

MRC.VER.01-02.10