Ultrasonic milk analyser

After 5th measurement completion automatically appears the menu:

Recal 1

Put sample water


Which reminds for 5-times water measurement. There is no need to enter
values after placing the sample-holder. Measurement starts directly.
After the 5

th

measurement appears the menu:

Recalibrated

Analyzer ready


This means that the calibration was completed successfully and the analyzer
is recalibrated for cow milk, marked as “Calibr 1”.
Switch off the power supply of the device and switch it on again.
The device is ready to work with the new calibration.

If during work the sample’s temperature exceeds the temperature range 15-
25° С the following message appears:

Temperature out of

range


Wait till the end of the measurement. When the message appears:

Put sample again


Put sample with temperature in the temperature range and continue the
measurement till completion of 5-time measurement.

Operation manual

MRC.VER.01-02.10