Ultrasonic milk analyser


5.5. Determination of total proteins content
We recommend using the following formulas:

А/ for cow milk

Protein = SNF * 0,367 (% )

Where
SNF

- solids-non-fat content in percentages (%),

0,367

– constant coefficient.


B/ for sheep milk

Protein = SNF * 0,475 (% )

Where
SNF

– solids-non-fat content in percentages (%),

0,475

- constant coefficient.