C. To select a Group, press the SELECT button when the Group you desire appears in the display (failure

to press the SELECT button within 3 seconds will activate the Group shown in the display and return the
S310 to its operational state).

D. Press the SETUP button until the "ULtr" annunciator appears in the display. Immediately press the

SELECT button to select the Ultra detector configuration program.

E. The "SET CAL" (calibration) annunciator now appears in the display. Press the count up (RANGE) and

count down (MODE) buttons to select the calibration constant (mV/W) value listed on the serial tag of
your Ultra detector. Press the SELECT button to enter the calibration constant value to memory.

F. The "SP" (speed) annunciator now appears in the display. Press the count up (RANGE) button to speed

up or count down (MODE) button to slow down the response time of the meter. Press the SELECT
button when the desired number appears in the display. We recommend initially setting the speed to 0.3.
Press the SELECT button to enter the speed setting to memory.

G. The "At" (attenuation) annunciator now appears in the display. If no attenuator is being used in

conjunction with the detector, the attenuation factor in the display must be set at 1.000 since this value is
a display multiplier. Attenuation multipliers from .0001 to 9999 can be entered into the displayed value.

The attenuation factors of optics can be entered into the H410 so the displayed value will automatically
compensate for the amount of attenuation. For example, assume a beam splitter is being used that
transmits 75% and reflects 25% of the beam. If the H410 is set up to measure the reflected beam the
attenuation could be set up as follows:

• An attenuation factor of 1 would display the value of the reflected beam.

• An attenuation factor of 3 would display the value of the transmitted beam.

• An attenuation factor of 4 would display the value of the source.

Use the count up (RANGE), count down (MODE) buttons to change the attenuation factor. Press the
SELECT button to enter the attenuation factor to memory.


H. The REMOTE annunciator now appears along with the digital interface previously selected; r232

(RS232) or IEEE (IEEE488). Press the SETUP button to change the interface between r232 (RS232) and
IEEE (IEEE488). Press the SELECT button when the desired interface appears in the display.

I. If RS232 has been selected, the "br" (baud rate) annunciator appears in the display along with the baud

rate previously selected. To change the baud rate press SETUP button until the desired baud rate appears
in the display. One of the following baud rates can be selected: 300, 1200, 2400, 9600, and 19,200.
When the preferred baud rate appears in the display press the SELECT button.

J. The "PA" (parity) annunciator now appears in the display. Press the SETUP button to select none, even,

or odd parity. Press the SELECT button to enter the parity choice to memory

K. The "HS" (handshake) annunciator now appears in the display. Press the SETUP button until your

choice of none (nOnE), on/off (onoF), or clear to send (CtS) appears in the display. Press the SELECT
button when your choice is displayed to enter the handshake to memory.

L. If IEEE488 has been selected, the "bA" annunciator appears with the default bus address. Bus addresses

from 0 to 30 may be selected by pressing the count up (RANGE) count down (MODE) buttons followed
by the SELECT button. Once the remote interface has been setup, the group annunciator will appear.
Press the SELECT button or simply wait a few seconds and the group will automatically be selected for
operation.