1. Calorimeter Watts Mode:



Compare voltage to the watt reading displayed on the
S310 to determine V/W constant to use when looking at
the analog output.




In power (watts) mode, the analog output is nominally 3 V full scale for each range. Therefore, to calculate watts
from the analog output voltage:

W = V/S

where:
V

=

analog

output

voltage

(V)

S = analog output sensitivity (V/W)

≈ 3 V/range (W)

(for example, on the 10 mW range, S

≈ 3 V/.010 W = 300 V/W)

2. Calorimeter Joules Mode:


Compare the peak voltage to the energy (joules)
reading displayed on the S310 to determine
V/J constant to use when looking at the analog output.





In the energy (joules) mode the analog output is an amplified voltage signal proportional to the voltage generated
by the calorimeter thermopile. To calculate energy (joules) from the analog output use the following equation.

J = 0.3(Vpk)(TC)

(S)(R)

where:

Vpk = peak voltage from the analog output

TC

= the calorimeter time constant from the calorimeter serial tag

S

= the calorimeter output sensitivity as follows:

0.5

V/W

for

25

mm

models

0.1667

V/W

for

50

mm

models and large aperture calorimeters

R

= indicator range multiplier as follows:

10.0

for

the

10

or

30

range

100.0

for

the

1

or

3

range

1,000.0

for

the

.1

or

.3

range

10,000.0 for

the

.01

or

.03

range