Scientech H410 Vector User Manual

Page 20

Advertising
background image

20

AC2500, AC25HD

60 seconds

AC2501, AC25HD, AC2504

60 seconds

AC5000, AC50HD

90 seconds

AC5001, AC50UV, AC5004

90 seconds

360401 with interface module

105 seconds

380401, 380402, 384UV5 with interface module 180 seconds
360801 with interface module

125 seconds

380801, 380802, 388UV5 with interface module 200 seconds

Note: Make sure the calorimeter delay is less than the time out setting you will make in section 10.


K. Press and hold the OFF/CANCEL button until the display blanks to save the settings.


L. Press the ON/SELECT button to turn the H410 back on.

6. To Select a Range:

Note: AUTO range is not available in the energy mode for calorimeters.
Note: You must save all settings by pressing and holding the OFF/CANCEL button until the

display blanks. If the H410 is turned off, by using the ON/OFF slide switch, before settings
are saved they will be lost.

A. Press the RANGE button. The analog scale (if active) will disappear and the available ranges will

cycle through the display. The ranges available for calorimeters are in the following table:

Model

Astral 25mm

Astral 50mm

Large Aperture 100mm

with PN10735 Interface

Module*

Large Aperture 200mm

with PN10747 Interface

Module**

Mode

Power Energy Power Energy Power Energy Power Energy



Range

10mW

100mW

1W

10W

AUTO

10mJ

100mJ

1J

10J

300mW

3W

30W

AUTO

300mJ

3J

30J

300mW

3W

30W

AUTO

300mJ

3J

30J

300mW

3W

30W

AUTO

300mJ

3J

30J

* With a PN10748 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 50 W or

150J.

** With a PN10769 – 10X attenuator, the actual power or energy is 10 times the displayed value up to 100 W or

300 J.

B. To select a range, press the ON/SELECT button when the range you desire appears in the display. The

selected range will be activated and the H410 will return to the last operational state.


C. Press and hold the OFF/CANCEL button until the display blanks to save the settings.


D. Press the ON/SELECT button to turn the H410 back on.


Note: When using PN10748 or PN10769, 10X attenuators, you may want to use an attenuation factor of

10 so the displayed reading on the S310 is correct. However the range designation may be
incorrect if the power or energy reading exceeds the upper limit of the range. See the Group
Settings section for details.

7. To Select a Measurement Mode:

The modes available for the calorimeters are: Energy (J) and Power (W).