Arge, Perture, Alorimeter – Scientech H410 Vector User Manual

Page 8: Pecifications, Stral, Hotodiode, Etector

Advertising
background image

8

L

ARGE

A

PERTURE

(100

MM

&

200

MM

)

C

ALORIMETER

S

PECIFICATIONS

:

Model

360401

380401

380402

384UV5

360801

380801

380802

384UV5

Type of Absorber

Surface

Volume

Volume

Volume

Surface

Volume*

Volume

Volume

Aperture Diameter

100mm

200mm

Minimum Beam Diameter
Spectral Response

.25 - 35µm .266 - 1.2µm

9 -11µm

.193 - .36µm .25 - 35µm .266 - 1.2µm

9 -11µm

.193 - .36µm

Maximum Average Power

50W with full illumination of absorbing surface

100W with full illumination of absorbing surface

Minimum Average Power

150mW

700mW

Noise Level

1.5mJ - mW

7mJ - mW

Maximum Power Density

200W/cm²

See Note 1

4W/cm²

Note 2

200W/cm²

See Note 3

4W/cm²

Note 2

Maximum Peak Power Density

1MW/cm²

See Note 4 100MW/cm² See Note 5

1MW/cm²

See Note 6 100MW/cm² See Note 5

Maximum Single Pulse Energy
Maximum Energy Density

Note 7

Note 8

4J/cm

2

Note 9

Note 7

Note 10

4J/cm

2

Note 9

Precision

< 1%

Accuracy

5%

Response Time

5 sec when connected to a Scientech Indicator in Watts Mode

Dimensions DxL - inches/cm

6.00 x 8.00/15.24 x 20.32

9.00 x 10.00/22.86 x 25.40

Weight - pounds/kgs

6/2.72

16.27/7.26

Indicator Compatibility

H410, H410D, S310, S310D

* This is a segmented absorber
Note 1: 380401

27W/cm

2

@ 1064 nm, 21W/cm

2

@ 532 nm, 7.7W/cm

2

@ 355 nm, 158mW/cm

2

@ 266nm

Note 2: 384UV5, 388UV5

50W/cm

2

@ 355nm

Note 3: 380801

13.5W/cm

2

@ 1064 nm, 10.5W/cm

2

@ 532 nm, 3.85W/cm

2

@ 355 nm, 79mW/cm

2

@ 266nm

Note 4: 380401

90GW/cm

2

@ 1064 nm, 71GW/cm

2

@ 532 nm, 27GW/cm

2

@ 355 nm, 530MW/cm

2

@ 266nm

Note 5: 384UV5, 388UV5

Repetitive pulses: 101MW/cm

2

@ 355nm

Single pulses: 3.5GW/cm

2

@ 355nm

Note 6: 380801

45GW/cm

2

@ 1064 nm, 35.5GW/cm

2

@ 532 nm, 13.5GW/cm

2

@ 355 nm, 265MW/cm

2

@ 266nm

Note 7: 360401, 360801

Max J/cm

2

= 1000 x (pulse width)

1/2

to a maximum of 200J/cm

2

Note 8: 380401

Repetitive pulses:
3.7J/cm

2

@ 1064nm, 2.9J/cm

2

@ 532nm, 1J/cm

2

@ 355nm, 20mJ/cm

2

@ 266nm

Single pulses:
7J/cm

2

@ 1064nm, 5.6J/cm

2

@ 532nm, 2.1J/cm

2

@ 355nm, 41mJ/cm

2

@ 266nm

Note 9: 384UV5, 388UV5

Repetitive pulses: 1.1J/cm

2

@ 355nm

Single pulses: 40J/cm2 @ 355nm

Note 10: 38-0801

Repetitive pulses:
1.85J/cm

2

@ 1064nm, 1.45J/cm

2

@ 532nm, 0.5J/cm

2

@ 355nm, 10mJ/cm

2

@ 266nm

Single pulses:
3.5J/cm

2

@ 1064nm, 2.8J/cm

2

@ 532nm, 1.05J/cm

2

@ 355nm, 20.5mJ/cm

2

@ 266nm

150J

300J

5cm

7.5cm

A

STRAL

P

HOTODIODE

D

ETECTOR

S

PECIFICATIONS

:

Model

AP30

AP30UV

Type of Absorber
Aperture Diameter
Spectral Response

400 nm - 1.1

µm

200 nm - 1.1

µm

Maximum Power

30 mW

2 mW

Minimum Power
Maximum Power Density

170 mW/cm

2

5 mW/cm

2

Noise Level
Accuracy
Response Time
Dimensions D x L - inches/cm
Weight - pounds/kgs
Indicator Compatibility

H410, H410D

± 5 %

1 second

2.5 x 1.1/6.35 x 2.79

0.25/0.114

Silicon

7.9 mm

100 nW

1 nW

Advertising