H410, Ector, Ndicator – Scientech H410 Vector User Manual

Page 5: Pecifications, Bsorption of, Bsorbing, Aterial, Wavelength

Advertising
background image

5

V

ECTOR

H410

I

NDICATOR

S

PECIFICATIONS

:

Model

H410

H410D

4digit LCD with Selectable

Analog Meter Movement

Full Scale Ranges with Astral 25mm Calorimeter
Full Scale Ranges with Astral 50mm Calorimeter
Full Scale Ranges with Vector Pyroelectric Detector
Full Scale Ranges with Photodiode Detector - Watts only
Maximum Repetition Rate with Calorimeter in Joules Mode
Maximum Repetition Rate with Calorimeter in Watts Mode
Maximum Repetition Rate for Collecting Data in
Statistics Mode with a Pyroelectric Detector
Response Time with Calorimeter in Joules Mode
Response Time with Calorimeter in Watts Mode
Operating Temperature
Power Requirement
AC Charger Input Requirements
Dimensions H x W x D - inches/cm
Weight - pounds/kgs

4 ea. Rechargeable Batteries

8.25 x 4.0 x 1.5/20.96 x 10.16 x 3.81

1.23/0.56

120 Volts, 60 Hz

± 10 % or 220 Volts, 50 Hz ± 10 %

300 pps

Calorimeter Dependent - 1 to 3 seconds

Calorimeter Dependent - 3 to 10 seconds

5

o

C to 40

o

C

3.000 m, 30.00 m, 300.0 m, 3.000, AUTO

30.00

µ, 300.0 µ, 3.000 m, 30.00 m, AUTO

Calorimeter Dependent - 1 pulse every 60 to 90 seconds

Unlimited

Display

4 Digit LCD

10.00 m, 100.0 m, 1.000, 10.00, AUTO (Watts only)

300.0 m, 3.000, 30.00, AUTO (Watts only)

A

BSORPTION OF

HD

A

BSORBING

M

ATERIAL

:

Warning: You must exercise caution when using HD detectors. They exhibit spectral reflection of

between 7% and 18%, of the input power, back out of the aperture. Please refer to
Figure 1 to determine the reflectance for the wavelength you are measuring. These
detectors should be treated as a partial mirror or any other type of reflective optic and
the appropriate caution level observed, especially at the CO

2

wavelength.

A B S O R P T I O N O F H D A B S O R B E R V s . W A V E L E N G T H

5 0

5 5

6 0

6 5

7 0

7 5

8 0

8 5

9 0

9 5

1 0 0

0

1

1 0

1 0 0

W A V E L E N G T H [ M I C R O M E T E R S ]

ABS

O

R

P

T

IO

N [

%

]

C A U T I O N : S P E C U L A R R E F L E C T I O N

Figure 1


Note: HD detectors might show “beam” marks on the absorbing surface. These marks are

characteristic of the material and do not affect the performance of the detector. Do not rub off or
remove the marks. Polishing or cleaning the absorbing surface might change the performance of
the pyroelectric detector.