A.2 Periodic Maintenance

The TSM Series Low Profile Scale is designed and
engineered to require very little maintenance. Follow
the basic guidelines below:

1. Periodically clean any debris from under and
around the scale. If debris such as broken wood from
pallets wedges between the scale platform and the
ground or a ramp, erratic weights will result.

2. Provide adequate clearance for free platform
movement. Make sure a ramp(s) is not touching the
scale. Remove any debris under and around the scale
platform.

3. Make sure the scale is level. Adjust the leveling
feet if required to make sure all four feet are in contact
with the floor.

4. Periodic Calibration Notes

a. The scale is shipped factory calibrated.
b. It is recommended you have your scale

calibrated once a year with Test Weights to
assure accuracy.

c. You should refer to your quality system

guidelines to determine your particular
calibration frequency.

d. If calibration is required, refer to the

indicator manual to determine the correct
calibration procedures.

e. For optimum calibration, load the scale with

Test Weights equal to 70%-80% of the
scales capacity.

f.

The scale may be calibrated with less Test
Weight. Try to use at least 40% of the
scales capacity if possible.