The mechanical vacuum control provides the vacuum in

the bottle in a range of 150 mbar (minimum switching off

pressure) to 500 mbar (maximum switching on pressure)

for aspiration. At the end of the operation, aspirate all liq�

uids out of the aspiration tubing and rinse with a cleaning

agent or disinfectant if necessary.

An integrated non-return valve avoids backflow from the

vacuum supply to the BVC basic / BVC basic G. This helps

avoid contamination of the BVC with media that may be

inside the vacuum supply.

When separating the device from the vacuum supply, the

non-return valve avoids venting of the collecting bottle.

The aspiration system doesn’t feature an air admittance

valve. If the collecting bottle is under vacuum, the existing

underpressure inside the bottle remains until the vacuum

is reduced (pressure rises) by further aspiration.

During operation

+

Max. ambient temperature: 40 °C

max.

40°C

NOTICE

• Use the system only with the

integrated hy-

drophobic filter to protect the vacuum sup�

ply from aspirated liquids and aerosols, and

to protect the environment/user from con�

tamination risk.