Cleaning and decontaminating

Autoclaving

The collecting bottle with bottle head and screw

cap, the quick coupling and the filter are des�

ignated for steam sterilization at 121°C and 2

bar absolute (1 bar overpressure). Time of ex�

posure according to DIN 58946 t

e

= 20 minutes.

Prior to autoclaving loosen or remove the

bottle head from the bottle.

It is the user´s responsibility to ensure effective

autoclaving.

121°C

2 bar absolute

The number of autoclaving cycles can be

marked on the plastic disc (Memory Disc) of the

filter (max. 20 autoclaving cycles according to

the manufacturer instruction).

Autoclaving

Maximum liquid level in collecting bottle:

80%. When working with low boiling liquids or

liquids that tend to foam, the maximum liquid

level may be less.

max.

80%

121°C

2 bar

absolute

max. 20 x

NOTICE

NOTICE

Over time, discoloration and material changes (e.g., re�

siliency, elasticity/tightness, cracking) due to repeated

steam sterilizations may occur.