Troubleshooting, Fault possible cause remedy – VACUUBRAND BVC basic G User Manual

Page 29

Advertising
background image

page 29 of 35

Fault

Possible cause

Remedy

❑ No aspiration.

➨ Filter clogged?

✔ Replace filter.

➨ Leak at the bottle head? ✔ Check seals, coupling

and blind cap. Screw in

or replace if necessary.

➨ No vacuum supply con�

nected?

✔ Connect vacuum supply.

➨ No vacuum while vacu�

um supply connected?

✔ Check vacuum supply.

Open lock valve in the

vacuum line if applica�

ble.

❑ Pressure control

device switches fre�

quently.

➨ Leak in the system?

✔ Check tubing, seals,

coupling, and blind cap.

Screw in or replace if nec�

essary.

➨ Filter clogged?

✔ Replace filter.

➨ Vacuum controller con�

taminated inside or

defective?

✔ Contact local distributor.

Troubleshooting

Advertising
This manual is related to the following products: