Vacuum connection (inlet) – VACUUBRAND PC 3012 NT VARIO Duo User Manual

Page 42

Advertising
background image

page 36 of 145

Vacuum connection (inlet)

Inlet: Small flange connection KF DN 25 with clamping

ring and centring ring.

Connect the vacuum line (e.g., metal hose (stainless steel)

KF DN 25) at the inlet of the pump.

• Reduce the transmission of vibration. Pre-

vent mechanical load due to rigid pipelines.

Insert elastic hoses or flexible elements

as couplings between the pump and rigid

pipes.

Note: Flexible elements will compress or

flatten when evacuated if not designed for

use under vacuum.

• Hose connections at the pump inlet must al-

ways be gas tight.

Use a suitable adapter to connect a vacuum

hose (caoutchouc) 15 mm ID (see “Accesso-

ries / Spare parts”, pg. 92), if necessary.

inlet

Use connecting hoses with large diameter and

keep them as short as possible to avoid flow

losses. Locate the pump as closely as possi-

ble to the application.

Secure hose connections at the pump ap-

propriately, e.g., with hose clamps, to protect

against accidental detachment.