Duration: Duration” determines the total process time since control

start.

The ”Duration” is adjustable between 1-1440 minutes (24 h) or can

be set to ”Off”. A preset ”Duration” (process time) has no effect, if the

process is stopped earlier by reaching the preset ”Minimum”.

The screen-shot shows the factory-set values.

Temporary switching from ”Auto mode” to ”Vac control” function

Press ”MODE” key. The controller switches to ”Vac control”. The cur-

rent pressure value is used and held as new set point.

The preset function of the controller does not change due to this tem-

porary switching. When pressing ”STOP” key, the controller is set

again to the ”Auto mode”.

50 mbar

00:07:45

Auto mode

When selecting ”Graphic” the display shows a pressure vs. time curve.

The timeline in the diagram adapts automatically to the process time.

Press the selection knob twice to return to the standard display.

Auto mode
Sensitivity normal

Speed

HI

Minimum

Off

Delay

Off

Duration

Off

- - - - - - Graphic - - - - - -

- - - - - - - Back - - - - - - -

00:00:00