During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

Maximum ambient temperature: 104 °F

(40 °C)

Check the maximum temperatures, if install-

ing the pump in a cabinet or a housing.

Make sure ventilation is adequate, especial-

ly if the ambient temperature is elevated.

max.

104°F

(40°C)

• If the pump is installed at an altitude of more

than 6500 ft (2000 m) above mean sea lev-

el, check compatibility with applicable safety

requirements, and adopt suitable measures.

There is a risk of the motor overheating due

to insufficient cooling.

max. 6500 ft

(2000 m)

• Check compatibility with the

maximally

permitted pressure at outlet and the max-

imum pressure difference between inlet

and outlet ports.

max.

16.0 psi

(1.1 bar)

absolute

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.