Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check carefully

the validity of this document with respect to his product. manual-no.: 999128 / 29/04/2014

Electronic equipment is never 100% fail-safe. This may lead to an indefinite status of

the equipment. Provide protective measures against malfunction and failure.

Operating the pump, stand still of the pump or operating the air admittance valve

must not lead to a critical dangerous situation under any circumstances.

Ensure that in case of failure the pump and the vacuum system always will turn into

a safe status.

In case of leaks in the manifold or diaphragm cracks pumped substances might

be released into the environment or into the pump housing. To reduce the risk of

leaks, ask for a diaphragm pump with additional safety diaphragm.

Obey especially notes on operation and use and maintenance.

Failure of the pump (e. g. by power failure) or connected components, parts of

the supply (e. g. coolant) or change of parameters (e. g. increase of pressure

in the coolant system) must not lead to a critical dangerous situation under any

circumstances.

Adopt suitable measures to prevent the release of dangerous, explosive, corro-

sive or polluting fluids.

If pumping

different substances, it is recommended to purge the pump with air

or inert gas prior to changing the pumped media in order to pump out residues

and to avoid reactions of the pumped substances with each other and with the

pump materials.

Take into consideration interactions and chemical reactions of the pumped

media. Ensure that the materials of the wetted parts are compatible with the

pumped substances, see section ”Technical data”.

The valves and the diaphragms are wear parts. Replace the diaphragms and

valves at the latest at 90% of the typical lifetime or immediately at higher noise

level. The typical lifetime of a diaphragm is 15000 operation hours when pumping

non-corrosive or non-condensable gases, if necessary provide an operation time

counter.

Check the overpressure safety valve between the pump stages (only pumps with

four cylinders) regularly and replace at the latest at 90% of the typical lifetime

(typical lifetime: 15000 operation hours).

Take adequate precautions to protect people from the effects of dangerous sub-

stances (chemicals, thermal decomposition products of fluoroelastomers), wear

appropriate safety-clothing and safety glasses.

Obey applicable regulations when disposing of chemicals. Take into considera-

tion that chemicals may be polluted.

Adopt suitable measures to prevent contamination of the pumped substances or

the environment.