catchpot
at the outlet

catchpot

at the inlet

Assemble catchpot at the inlet and at the outlet using joint

clips (fig.: MZ 2C EX + AK + EK)

coolant outlet
(hose nozzle 6 mm)

outlet

(gas!; hose nozzle 10 mm)

coolant inlet
(hose nozzle 6 mm)

Assemble hose nozzles for cooling water inlet and outlet at

the vapour condenser.

MZ 2C EX + AK + EK / MZ 2C EX + IK + EK

Attention: Glass parts (flasks and condensers) of pumping units (pumps

with AK (IK) and EK) have no protection against splinters (against implo-

sion) or against leakage in case of mechanical damage or hits from out-

side.

The user must ensure an appropriate protection against splinters / implo-

sion or leakage prior to the first use.

Pumps with two motors

Ensure that in case of pumps with two motors both motors are switched on at the

same time, if not possible switch on the motor of the pump at the outlet first.

Connect the inert gas connections at the crankcase not serial but parallel.

Assemble hose nozzles for cooling water inlet and outlet at

the vapour condenser.

1

2

Assemble catchpot at the inlet (1) using joint clips.

The collecting flask at the outlet (2) is already assembled.

MD 4C EX + AK + EK / MV 10C EX + AK + EK

outlet

(gas!; hose nozzle 10 mm)

coolant outlet
(hose nozzle 6 mm)

coolant inlet
(hose nozzle 6 mm)