Unscrew the housing plate from the pump.

Position the pump on the motor.

Unscrew the locking plate at the inert gas connection and

remove hose.

Use open-ended wrench to unscrew the inert gas fitting at

the housing cover.

Tools required (metric):

- Phillips screw driver size 2

- Open-ended wrench size 10/15/17

- Open-ended wrench size 19, 4.5 mm thick

- Torque wrench size 19

- Hex key size 5

- Diaphragm key SW 66 (cat. no.: 636554)

- Open-ended wrench size 19 with torque indicator

Cleaning and inspecting the pump heads

Unscrew the union nut at the elbow fitting of the pump

head and remove hose.

Turn the fittings as little as possible and do not remove the

fittings from the pump head.

Through reassembly a leak may result.