Notes on operation

For operation it is necessary to install valves and/or vacuum pumps.

Max. permitted pressure at pressure transducer: 2 bar (abso-
lute).
☞ The display flashes at a pressure higher than 1060 mbar.

☞ Inside a vacuum system where evaporation occurs, the

vacuum is not uniform. This effects the value of the meas-
ured pressure and such the controlling. Therefore carefully
choose the position where to connect the pressure trans-
ducer.

☞ Condensate and deposits at the pressure transducer falsify

the measurement result.

☞ If residues occur or when working with aggressive or con-

densable substances, install a gas washing bottle in front of
the pressure transducer.

Position controller in such a way, that condensate can

not flow into the pressure transducer.

☞ If necessary, clean pressure transducer, see section ”Clean-

ing the pressure transducer”.

☞ Setting of interface parameters, see ”Interface parameters”.

☞ Presettings at controller, see ”Factory-set modes and menu structure”.

☞ Operating the controller, see ”Working with the controller”.

vacuum connec-
tion

Assignment of pins

6

7

8

3

2

1

5

4

1

cooling water valve

2

+ 24V

3

isolation valve

4

external error -

5

external error +

6

mass (PWM -)

7

PWM + (speed control signal for VARIO pumps)

8

+12 V

Note: Texts written in Courier font mirror the display of the LCD of the controller.