Working with the controller – VACUUBRAND VNC 1 User Manual

Page 18

Advertising
background image

Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999118 / 16/12/2009

page 18 of 35

Program menu in basic mode pControl

pControl

p Set

pControl
Δp

pControl

t Off

p Set

25mbar

Δp
Auto

t Off

off

1013mbar

Start

Esc

Working with the controller

Factory-set mode ”Vacuum control” (pControl)

Control without isolation valve:

Control of a vacuum pump by direct switching of the pump. Preset time delay is only valid for an
optional cooling water valve.

Control with isolation valve:

Control of a vacuum pump via an isolation valve. Preset time delay is valid for the pump to reduce
condensation inside the pump considerably. The cooling water valve is open during vacuum control and
during time delay.

For synchronous switching of the pump and the isolation valve:
Configure the time delay in the device menu: t Off: off.

Operation of a speed controlled (VARIO) vacuum pump:

Connect the pump using a Y-type adapter cable (see accessories) to the controller. The external error
input must not be configured (factory-set: ext.Err: off). Switch on the controller and start regulation.
After approx. 1 second the VARIO pump is identified automatically.

☞ Use keys ▲ or ▼ in the program menu to select an item.

☞ Confirm selection with key ”Enter”.

☞ Adjust parameters with keys ▲ or ▼.

☞ Confirm with key ”Enter” or return to original setting pressing ”Esc”.